A low-frequency noise of about 0.5 to 0.6 Hz called baseline wander. It can be eliminated using a high-pass filter with a cut-off frequency of 0.5 to 0.6 Hz. By applying a notch filter with a 50 or 60 Hz cut-off frequency, powerline interference (50 or 60 Hz noise from the mains supply) can be eliminated.
Patient movement is a common cause of choppy interference throughout the trace; even shivering or chewing gum can result in somatic tremor artifact. Try providing a warm blanket to your patient if they are feeling cold or frightened.
Incomplete baseline artifact is brought by by. Patient cable breaks when the metal tip of the lad wire separates from the alligato clip. for baseline artifact interruption correction. Change the patient cable, or fasten the metal lead wire tip to the alligator clip.